Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Beckman RMS225: CAL ADJ /5520 DATE: 27-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 10 NUMBER OF LINES: 105 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # none # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: NONE 1.004 DISP Temperature: 18 - 28 degC 1.004 DISP Humidity: <80% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP [32] WARNING 1.006 DISP 1.006 DISP To prevent a shock hazard, remove all test leads 1.006 DISP from the instrument and turn the instrument OFF 1.006 DISP before disassembling the RMS225. 1.006 DISP 1.007 HEAD DISASSEMBLING INSTRUCTIONS 1.008 DISP 1. Remove the unit from the holster(if necessary) 1.008 DISP 1.008 DISP 2. Place the unit face down and remove the 1.008 DISP [32] four screws securing the rear cover. 1.008 DISP 1.008 DISP 3. Remove the rear cover. 1.009 HEAD LCD ADJUSTMENT 1.010 DISP Set the RMS225 FUNCTION to V(DC). 1.010 DISP 1.010 DISP Short the V-OHM input to the COM input. 1.011 OPBR The LCD ADJUSTMENT is for the Self-resetting 1.011 OPBR fuse models ONLY! 1.011 OPBR 1.011 OPBR Does this unit have the self-resetting fuse? 1.012 JMPF 1.015 1.013 DISP ADJUST - R26(lower left) for sharp LCD segments 1.013 DISP without ghosting. 1.014 RSLT =R26 was ADJUSTED for a sharp LCD display. 1.015 HEAD {BASIC (DC)} 1.016 DISP Connect the 5520A to the UUT as follows: 1.016 DISP [32] 5520A NORMAL HI to UUT V/OHM 1.016 DISP [32] 5520A NORMAL LO to UUT COM 1.017 5520 2.500V S 2W 1.018 DISP Adjust R6 (DC Cal) for EXACTLY 2.500 on display. 1.019 5520 -2.500V 0.002U 2W 2.001 JMP 2.004 PASS 2.002 OPBR Negative reading is not -2.500 +/-0.002V. 2.002 OPBR Would you like to repeat the adjustment? 2.003 JMPT 1.017 2.004 HEAD {AC VOLTAGE} 2.005 DISP Set the RMS225 FUNCTION to V(AC). 2.006 5520 0.000V 0.008U 2W 3.001 5520 2.500V 60H SI S 2W 3.002 DISP Adjust R15 (AC Full Scale) for EXACTLY 2.500 +/-.002. 3.003 5520 2.500V 10kH SI S 2W 3.004 DISP Adjust C28 (10kHz) for EXACTLY 2.510 +/-.004. 3.005 5520 2.500V 0.014U 10kH SI 2W 4.001 JMP 4.004 PASS 4.002 OPBR UUT does not read 2.500 +/-0.014V. 4.002 OPBR Would you like to repeat the adjustment? 4.003 JMPT 3.003 4.004 5520 25.00V 10kH SI S 2W 4.005 DISP Adjust C29 (10kHz) for EXACTLY 25.10 +/-.04. 4.006 5520 25.00V 0.14U 10kH SI 2W 5.001 JMP 5.004 PASS 5.002 OPBR UUT does not read 25.00 +/-0.14V. 5.002 OPBR Would you like to repeat the adjustment? 5.003 JMPT 4.004 5.004 HEAD {AC FREQUENCY RESPONSE} 5.005 5520 0.800V 0.02U 1kH SI 2W 6.001 5520 0.800V 0.02U 10kH SI 2W 7.001 5520 2.500V 0.046U 60H SI 2W 8.001 5520 2.500V 0.046U 10kH SI 2W 9.001 5520 25.00V 0.46U 10kH SI 2W 10.001 HEAD ASSEMBLY INSTRUCTIONS 10.002 DISP Disconnect ALL connections and turn unit OFF. 10.002 DISP 10.002 DISP Replace the rear cover and tighten the four screws. 10.003 END